AC/DC Wide input voltage Low power consumption Small size Full protections High reliability Medical Power Module ADM360K240S001A

Features:

● Wide input voltage: the input working voltage range is 100-240VAC

● Low power consumption: No-load <0.075W

● High Energy Efficiency: Six levels of energy consumption, power efficiency up to 90%

● High reliability: Comply with EN60601-1 CLASS Ⅱ safety level and 2×MOPP insulation protection level, and pass CE

● Flame retardant insulation:UL94V-0 flame retardant heat resistant material

● Protection types:short circuit protection, over current protection, over voltage protection, and self-recovery

● Small size:65.8*44.5*31.49mm, it can be put into products with high volume requirements


Specification

Product Tags

Specification

03
Input Parameters
Voltage Range 100~240 VAC
Frequency Range 50~60Hz
Input Current (Max.) 0.7A @ 115VAC,0.5A @ 230VAC
Efficiency (Typ.) 90%
Standby Consumption 0.075W
Impulse Current (Typ.) 10A @ 115VAC,23A @ 230VAC
Leakage Current (Typ.) 0.1mA @ 264VAC, 60Hz
Output Paremeters
Output Voltage 24VDC
Voltage Tolerance ±5%
Output Current 1500mA
Rated Power (Max.) 36W
Line Regulation ±0.5% at full load
Max. Capacitive Load 2000uF
Load Regulation ±1%
Ripple & Noise 50mV (Typ.), 100mV (Max.)@20MHz
Frequency (Typ.) 65kHz
Hold up Time (Typ.) 15mS @ 115VAC80mS @ 230VAC
Environment
Operating Temperature -40 ~ +70℃
Storage Temperature -40 ~ +85℃
Storage Humidity (Max.) 95%RH (Max.)
Power Derating 2.7%/℃ @ -40 ~ -25℃

2.4%/℃ @ +50 ~ +70℃,12VDC

2.7%/℃ @ +55 ~ +70℃

Temperature Coefficient ±0.02%/℃
Soldering Temperature 260±5℃ @ Wave Soldering, 5~10s

360±10℃ @ Manual Soldering, 3-5s

Protection Function
Short Circuit Long term short circuit, auto recovery.
Over Current ≥130%IO auto recovery
Over Voltage ≤27VDC @ 12V output, ≤40VDC @ 24V output
IEC Safety Class CLASSII
Electric Shock Protection 2×MOPP @ primary to secondary
Reliability
MTBF ≥300000H @ 25℃
Safety Standards / Directives
Medical safety EN60601-1
CE Compliant
Isolation Voltage (Min.) Input to Output 4000VAC @ 1 minute test and the leakage current is smaller than 5mA.
EMC
 

 

 

EMI

 

 

 

 

Parameter Standard Test Level / Note
Conducted emission EN55011(CISPR11)/EN55032(CISPR32) CLASS B
Radiated emission EN55011(CISPR11)/EN55032(CISPR32) CLASS B
Voltage flicker EN61000-3-2 -
Harmonic current EN61000-3-2 -
EMS

 

 

 

 

 

 

Parameter Standard Test Level / Note
Electrostatic Discharge IEC/EN61000-4-2 ±8KV/Contact

±2、4、8、15KV/Air

Radiate Susceptibility IEC/EN61000-4-3 10V/m
Electrical Fast Transient burst IEC/EN61000-4-4 ±2 KV
Surge IEC/EN61000-4-5 ±0.5/±1 KV

Professional/Family medicine

Conducted Susceptibility IEC/EN61000-4-6 3Vm/0.15MHz-80MHz 6Vm

(Within15m band)

0.15MHz-80MHz

80%AM,1kHz

Voltage Dips and Interruption IEC/EN61000-4-11 0%UT 0.5cycle

0°,45°,90°,135°,180°,270°,315°。

0%UT 1cycle

70%UT 25/30 cycle 0°

0%UT 250/300 cycle

 

 

Mechanical Specification
Dimension 65.79*44.54*31.49mm
Enclosure Material Black flame retardant and heat resistant plastics(UL94V-0)
Cooling mode Natural air cooling
Application

1


  • Previous:
  • Next:

  • Related Products